Description
2.2.1 Scanning Mode: Convex and linear array scanning.
2.2.2 Display Mode: B, 2B, 4B, M, B/M, ZOOM B
2.2.3 Probe (Basic Configuration)
Super broadband high-density multi-frequency convex probe
2.2.4 Focusing Method:
a) Transmitting focusing method: 1 ~ 4 transmitting focus(es) for selection transmitting focus(es) for selection, 16 focusing positions for selection
b) Receiving focusing method: continuous dynamic focusing
2.2.5 Gray Scale:256
2.2.6 Beam Forming Method:
Digital beam forming, continuous dynamic focusing, dynamic aperture, dynamic, apodization
2.2.7 Display depth: Max 25.2cm
2.2.8 Probe technology:
a) Broadband probe
a) Frequency range: 2.5MHz ~9.0MHz (which can be extended to 2.0MHz~12MHz)
2.2.9 Imaging Parameter Adjustment:
B gain, M gain, 8-step TGC (Time Gain Compensation, also called “STC”), acoustic
power, focal point number, dynamic range, Smo/Edg, image persistence (frame
correlation), line density and SRT(Speckle Reduction Technology).
2.2.10 Live Zoom: 4 magnifications with the max. of ×4.0, position selectable
2.2.11 Display Parameter Adjustment:
Display depth, display angle (for convex probes), display width (for linear probes), image orientation (left/right reverse and 90° rotation), image polarity (positive / negative), mage grayscale curve
2.2.12 Full Screen Sweep Speed In M Mode (s/f): 1.25, 2.5, 5.0 and 10.0 for selection
2.2.13 Cineloop: Current stored image cineloop, and hard disk cineloop; ≥512 frames, and the replay speed is selectable.
2.2.14 Image Storage: Hard disk, USB; Temporal image storage and recall: ≥32 frames
2.2.15 Body Mark: ≥90 types for selection, with probe marks
2.2.16 Annotation: not less than 300 system predefined annotations
2.2.17 System preset
2.2.18 Image Screen Display:
a) Display all parameters related with diagnosis: system model, hospital name, patient name, patient ID, date, time and week, probe model exam type, and probe orientation;
b) Display all the parameters related to imaging: acoustic output power(% or dB), depth, frame rate, imaging angle/ width, focus mark and range, gray scale, probe operating frequency, TGC curve, depth scale, gain, Smo/Edg, frame correlation, grayscale curve, dynamic range, zoom rate, line density and SRT.
2.2.19 Storage and Record:
a) Hard drive, in which images can be stored in the formats of Jpg/Bmp or DICOM file and cineloop files in the format of Cine or Avi.
b) USB connection devices (Option)
c) transmit and receive images and patient data through DICOM3.0 function (Option)
d) Printer (Option)
2.2.20 Measurement and Calculation:
a) General measurement: B mode: Distance, area, circumference, volume, angle, histogram; M mode: Distance, time, heart rate and slope
b) Special measurement and calculation: for abdomen, cardiology, OB/ GYN, urology, small parts, peripheral vascular and orthopedics
2.2.21 Video Port: composite video out, PAL/ NTSC format for selection
2.2.22 Monitor: High resolution and non-interlaced monitor
2.2.23 Probe Sockets: 2 pcs
2.2.24 Power Supply:
a) Power voltage: 100V-240V~,tolerance±10%, or 230±23V~ for EU countries
b) Power frequency: 50Hz±1Hz or 60Hz±1Hz
c) Power input: 250 VA
2.2.25 Physical Dimension and Weight:
a) Physical dimension (LxWxH): 415mm×320mm×310mm
b) Weight: approx. 12kg
2.2.26 Operation: continuous operation
Reviews